GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 93
Image Generator
Select the type (CT/MR/PACS) from the listbox. If required for documentation, enter the
manufacturer and model name of the Image Generator.
Connection
Check the applicable connection status. In most cases, it will be a connection to a PACS
workstation.
Visual Test button
Starts the visual test part of the QA test. Follow the instruction in the QA Visual Test
Verification dialog box and visually verify the screen. Depending on selected QA test
standard, a series of QA test images will be displayed sequentially for your visual assessment.
Click Passed, if there are no problems and Failed if you recognize any problem or Abort to
exit the test.
Figure 74: QA Visual Test dialog
The process returns to the QA Test dialog, as shown in Figure 71 when all the visual tests are
done.
Measurement Test button
For measurement tests, use Measurement Test by clicking on this button. The Sensor
Contact Position Guide (Figure 57) will be displayed. Please position an external sensor on
the screen. The measurements will start when you click the Continue button.
NOTE: If an external sensor is not connected to the PC, the Measurement Test button is
inactive. Click the OK button once, then connect an external sensor to the PC and then - after
recognition of the sensor in Sensor Setup - please perform the QA Test again.
Display Category
The category is determined by required minimum luminance levels (measured in cd/m
2
).
These are fixed in the AAPM or ACR AAPM SIIM or JESRA standard.