GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 55
Figure 32: Status screen on target display calibration
Once a calibration is completed, a Calibration Report dialog box is shown on each target
display, as shown in Figure 33. See 7.2.3 Calibration Reports (page 56).
Figure 33: Calibration Report dialog box
NOTE:
- When the target curve Native was selected, a grayscale calibration will not be performed.
Only the native luminance response curve of the display will be measured. In addition,
calculation for the conformance test will be performed when “DICOM GSDF” or “L*” custom
curve is selected. For more information about setting the target curve, refer to 7.2.5
Calibration Setup (page 64).