GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 54
7.2. Calibration
Click Calibration in the Display Maintenance screen to display Figure 31. The Calibration
screen is shown.
Figure 31: Calibration screen
7.2.1. Perform a Calibration
Select the display to select the check box at the top of the display icons (Multiple displays
can be selected.). And then click Start Calibration at the bottom right of this screen, the
calibration will start for the selected display(s). Please follow the instructions on the screen.
If both display sensors and external sensors have been set for various displays, calibration
will start with those displays for which display sensors are set and then will continue with
displays for which external sensors are set.
Calibration is executed simultaneously with multiple display sensors, except with MD212G3,
MD215MG, MD211G5 and MD302C6 models, where the calibration is done sequentially.
Using external sensors, the calibration executed in ascending order of display numbers.
The calibration process includes:
White Luminance Calibration
Grayscale Calibration
Conformance Tests
The calibration status is displayed on each target display, as shown in Figure 32. When the
Overall Progress bar reaches 100%, the calibration will be complete. A running calibration
may be interrupted by the Cancel button at any time.