GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 14
2. Checking System Dependencies
2.1. External Sensors
External sensor drivers are included with the GammaCompMD QA Client. Please install
GammaCompMD QA Client before connecting any external sensor to the system. If
multiple external sensors are connected simultaneously, they will not be correctly
identified. Please connect only one external sensor.
2.2. Attaching MD-N2M5B Sensors and External Sensors
The NEC MD-N2M5B external sensor can be used to perform automated calibrations
on some display models. Some MD-N2M5B sensors and external sensors cannot be
stopped by the operating system (the [Safely Remove Hardware] icon is not shown in
the taskbar). To remove a sensor that does not have this icon, only remove it after
checking that the sensor is not in use. It is recommended that the sensor be removed
after stopping the GammaCompMD QA Client.
2.3. Updating GammaCompMD QA Client with MD-N2M5B sensor(s)
If updating from GammaCompMD QA Client Version 4.0.10, the settings of MD-N2M5B
sensor will be discarded. Set it up again after upgrading. For more information about
sensor settings, refer to 7.2.6 Sensor Setup (page 77).
2.4. Using the CA-210 Color Analyzer
If connecting with serial communication, set the baud rate for the sensor unit to
9600bps. The sensor will not be detected if it is set at another baud rate. Settings are
not necessary when connecting with USB communications. Also, special modes set at
the CA-210 (MEAS or 0-CAL) will not be recognized by GammaCompMD QA Client.
Please follow the instructions displayed at the start of calibration and startup to
properly set the mode. You cannot use a CA-210 which supports two or more
measuring probes. Only one probe connection is supported. Please refer to detailed
instructions how to use this instrument in the CA-210 user manual.
2.5. Using the IBA LXplus instrument
For ambient light measurement, the optional lux sensor needs to be attached to the
LXplus instrument. Please turn the LXplus power to OFF when attaching / detaching the
Lux sensor. At this time, be careful to not pull out the USB cable when performing USB
communications with a USB - serial conversion adapter. Please refer to detailed
instructions how to use this instrument in the LXplus user manual.