GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 161
No
Occurrence/Error
Solution
18
An error dialog appears stating
“Calibration was stopped due
to an error.”
Check whether sensor and display are
connected.
Calibration may not be possible with the
selected calibration target values.
Open the Calibration Setup dialog box, check
the selected calibration target values (refer to
7.2.5 Calibration Setup, page 64) and start
calibration again.
Check that the correct preferred sensor is
selected in Sensor Setup. Open the Sensor
Setup dialog box, check the Preferred Sensor
(refer to 7.2.6 Sensor Setup, page 77) and start
calibration again.
Close any other application which may use
the sensor.
19
An error dialog appears stating
“Calibration can't be
performed with the current
external sensor.
Please change the White
Luminance Calibration Mode or
use a color sensor."
An external sensor that supports color is
required, if the white luminance calibration
mode is different from Native or No Change.
Change the white luminance calibration
(7.2.5 Calibration Setup, page 64), or change to
an external sensor which supports color (7.2.6
Sensor Setup, page 77) and try again.
20
An error dialog appears stating
“Conformance test was
stopped due to an error.”
Check whether the sensor and the display are
connected.
Close any other application which may use
the sensor.
21
An error dialog appears stating
“Uniformity test was stopped
due to an error.”
Check that the external sensor
is connected and close any
other applications using the
sensor.
Check whether the sensor and the display are
connected.
Close any other application which may use
the sensor.