GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 69
Display Function The display function can be selected from the following four types
Native
When Native is selected as the target curve, a grayscale calibration is not performed. The
luminance characteristics of the display are only measured.
Gamma Curve Value
Fixed Gamma correction values from 1.00 to 3.00 may be selected.
DICOM GSDF
Default setting for DICOM Grayscale Standard Display Function calibration (DICOM Standard,
Part 14). Values from 0.50 to 1.00 may set when the Use Scale Factor checkbox is checked.
Custom Curve
When the Edit button is clicked, the Custom Gamma Curve dialog box, as shown in Figure 45,
is displayed. This shows a graphical representation of the display’s luminance characteristics.
Figure 45: Custom Gamma Curve dialog box
buttons
After dragging the mouse to specify a range on the graph in a rectangle, these
buttons can be used to zoom in or out on the curve.
L* and Log Linear buttons
Select pre-defined custom curves with these buttons.
Note: The NTSC Video, SMPTE-240M as well as the Import button are disabled for display
model which do not support this feature.