NEC GammaCompMD QA Manual de usuario Pagina 160

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 182
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 159
GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 160
No
Occurrence/Error
Solution
13
An error dialog appears stating
“Calibration parameters could
not be saved.”
The import files were not found for the custom
gamma curve. In Calibration Setup, go to the
Grayscale Function tab Custom Curve -> Edit
-> Custom Gamma Curve -> Import… and check
that the file to be imported exists and try again.
Refer to 7.2.5 Calibration Setup (page 64).
14
An error dialog appears stating
“Set first point to zero.”
The curve data being imported in the custom
gamma curve dialog is not suitable.
Make sure that the data starts from zero and
the figures do not decrease.
15
An error dialog appears stating
“Irregular increase in curve file
figures”.
The curve data being imported in the custom
gamma curve dialog is not suitable.
Make sure that the data starts from zero and
the figures do not decrease.
16
An error dialog appears stating
“White level measurement was
stopped due to an error.”
Check whether sensor and display are
connected.
Close any other application which may use
the sensor.
17
An error dialog appears stating
“Black level measurement was
stopped due to an error.”
Check whether sensor and display are
connected.
Close any other application which may use
the sensor.
Vista de pagina 159
1 2 ... 155 156 157 158 159 160 161 162 163 164 165 ... 181 182

Comentarios a estos manuales

Sin comentarios