GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 82
Sensor selection rules External sensors cannot be used for all calibration operations. For
scheduled display calibrations or remote calibrations and DICOM
conformance tests, only a display sensor is appropriate.
On the other hand, some external sensor models are unable to be
used for Display Quality Assurance (QA), depending on regional /
national regulations.
NOTE: If an external sensor was selected for a display with Display Sensor, but this external
sensor is not connected during a measurement or calibration activity, GammaCompMD QA
will automatically change to the display sensor, without notification. Only the respective
reports will reveal this change.
Using sensor models without color measurement capability
Typical instruments used for Quality Assurance, i.e. LXplus, LXcan
or Luxi/Solo Light do not support color measurements. White
luminance calibration modes, other than Native and No Change,
as well as a Reference Calibration of a color display front sensor
are not supported. Please refer to 7.2.5 Calibration Setup (page
64) regarding the White Luminance Calibration mode and
Reference Calibration button in this chapter regarding reference
calibration.